Eyecryl Sert EDOF – Preloaded IOL

Advanced Monofocal Hydrophobic Aspheric EDOF Foldable Intraocular Lens (Natural Yellow)

Eyecryl Sert Hydrophobic Aspheric EDOF IOL (Natural Yellow) IOLs are intended to be implanted into the capsular bag in the posterior chamber of the eye for the visual correction of aphakia secondary to the removal of the crystalline lens in adult patients with cataracts.

FEATURES

Aspheric Surface Design: The aspheric surface design of Eyecryl Sert is meticulously crafted to mimic the natural curvature of the eye, ensuring optimal light refraction. By reducing spherical aberrations, this design enhances image quality and contrast sensitivity, especially in low-light conditions. Patients experience sharper vision with improved clarity and reduced distortion.
Continuous Vision: Eyecryl Sert offers a seamless visual experience by providing continuous vision from distance to intermediate ranges. Unlike traditional monofocal lenses that offer clear vision at only one distance, Eyecryl Sert allows patients to see clearly across multiple distances without noticeable jumps or interruptions. This continuous vision facilitates daily activities such as reading, working on computers, and engaging in hobbies without the hassle of switching between different pairs of glasses.
Micro Incision Compatibility: Eyecryl Sert is compatible with Microincision Cataract Surgery, featuring a 2.2 mm incision size. This compatibility reduces surgical trauma, speeds up the recovery, and minimizes induced astigmatism, leading to better surgical outcomes.
Fast Neural Adaptation: Engineered for rapid neural adaptation, Eyecryl Sert enables patients to adjust quickly to their new visual environment post-surgery, ensuring a smoother transition and faster restoration of functional vision.

  • A new Advanced Hydrophobic EDOF IOL with Aspheric surface
  • Continuous vision from distance to intermediate
  • Spectacle independence for Intermediate Vision
  • An extended range of vision: >1.75 D at spectacle plane with Visual Acuity 0.2 LogMAR or better
  • Intermediate vision up to 57 cm
  • MICS (2.2 mm) compatible to reduce SIA
  • Designed for fast neural adaptation
  • Minimal glare and haloes
  • ABBE no. 47

OPERATIONS

Eyecryl Sert is designed for use in cataract surgery, where it replaces the clouded natural lens with a clear artificial one. Eyecryl Sert intraocular lens (IOL) is implanted into the capsular bag of the eye, restoring clear vision for individuals affected by cataracts. Eyecryl Sert’s compatibility with microincision cataract surgery techniques further enhances its utility, allowing for smaller incisions, faster recovery times, and reduced induced astigmatism.

ADVANTAGES

Enhanced Visual Quality: The aspheric surface design and minimized aberrations provided by Eyecryl Sert result in enhanced visual clarity, allowing patients to enjoy sharper and more precise vision post-operatively.
Predictable Refractive Results: Eyecryl Sert delivers predictable refractive outcomes through precise biometric calculations and a comprehensive range of available powers and toric options. Surgeons can rely on accurate preoperative measurements and calculations to select the appropriate lens power
Versatile Implantation: With a broad diopter range and toric options, Eyecryl Sert can address a wide range of patients’ refractive needs, offering versatility in surgical planning and customization.
High Patient Satisfaction: Post-operative data demonstrate high levels of patient satisfaction with Eyecryl Sert, including spectacle independence and satisfaction with visual quality and overall outcomes.

TECHNICAL SPECIFICATIONS

MODEL PLHFD6, PLHFD6T
MATERIAL Hydrophobic Acrylic Containing Natural Yellow Chromophore
OPTIC TYPE Single Piece, 360° Square Edge with Toric Aspheric Optic
OPTIC SIZE 6.00 mm
OVERALL SIZE 13.00 mm
ANGULATION 0
ACD 5.28
REFRACTIVE INDEX 1.524
RECOMMENDED ULTRASOUND A-CONSTANT SRK-T 119,0
RECOMMENDED OPTICAL A-CONSTANTS
  • HAIGIS: a0:1.574, a1:0.400, a2:0.100
  • HOFFER Q ACD: 5,78
  • HOLLADAY 1 SF: 1,99
  • HOLLADAY 2 SF: 2,02
  • Holladay 2 ACD: 5,78
  • SRK-T: 119,40
  • SRK-II: 119,90
  • Barrett: 2,09
DIOPTER RANGE +7.0 D to +30.0 D (with 0.5 D step) & 0.00 D to +6.5 D and +30.5 D to +40.0 D will be available on customization on case to case basis as per manufacturing possibility”
CYLINDER RANGE 1.0 D to 1.5 D (with 0.5 D step), 1.5 to 6.0 D (with 0.75 D step) – (For Toric Version).
IMPLANTATION SITE Capsular Bag
STERILIZATION EO
DELIVERY SYSTEM Pre-loaded Delivery System
SHELF LIFE 3 years from date of manufacture

CYLINDRICAL POWER OPTIONS & NEAR ADDITION

Model Near Addition Cylinder Power At IOL Plane Cylinder Power At Corneal Plane Recommended Range of corneal astigmatic correction*
PLHFD6T CYL D: 1.00 D 1.00 D 0.68 D 0.25 D – 0.86 D
PLHFD6T CYL D: 1.50 D 1.50 D 1.03 D 0.87 D – 1.25 D
PLHFD6T CYL D: 2.25 D 2.25 D 1.54 D 1.26 D – 1.75 D
PLHFD6T CYL D: 3.00 D 3.00 D 2.05 D 1.76 D – 2.25 D
PLHFD6T CYL D: 3.75 D 3.75 D 2.57 D 2.26 D – 2.75 D
PLHFD6T CYL D: 4.50 D 4.50 D 3.08 D 2.76 D – 3.25 D
PLHFD6T CYL D: 5.25 D 5.25 D 3.60 D 3.26 D – 3.75 D
PLHFD6T CYL D: 6.00 D 6.00 D 4.11 D 3.76 D and above